| ICHOR II Analyzer |
| Cat. No. |
Item |
| 5880 |
ICHOR II Analyzer |
| 5884 |
ICHOR II Diluent Solution (20L) |
| 5885 |
ICHOR II Rinse Solution (5L) |
| 5886 |
ICHOR II Lyse Reagent (500 mL) |
The ICHOR II is a lower-cost impedence cell counter for use with Plateletworks and hematology testing at the point of care. For ease of use, ICHOR II offers touch screen capabilities and cap piercing. ICHOR II also supports optional use of PS/2 keyboard, barcode scanner or printer. ICHOR II utilizes reagents bottles with reagent inventories managed by the instrument.
ICHOR II Specifications:
Tests: Hemoglobin, Hematocrit, RBC Count, RBC Indices (MCV, MCH, MCHC), WBC Count, Platelet Count, MVP, Whole Blood Platelet Aggregation with ADP, Collagen, Arachidonic Acid
Reagent Delivery System: External reagent and waste container bottles with reagent inventories managed by the instrument.
Linearity: RBC (x1012 cells/µL): 0.2 to 8.0 ±0.05 (or ±5%)
WBC (x109 cells/µL): 0.3 to 99.9 ±0.3 (or ±5%)
Hemoglobin (g/dL): 10 to 250 ±2 (or ±3%)
Platelets (x109 cells/µL): 10 to 999 ±10 (or ±10%)
Sample: Closed-tube sampling requires only 13 µL of whole blood
Throughput: 60 seconds per sample; 60 samples per hour
Operating Mode: Sequential, single channel analysis
Operating Principle: Electronic impedance measurement
Quality Control Features: Integrated QC program provides LJ, mean, SD, CV% analysis with graphing
User Interface: 23-key alphanumeric keypad (PS/2 keyboard optional)
Display: 800 x 600 character color LCD
Printer: Optional
Memory: Stores maximum of 35,000 sample results
Power Requirements: 100 to 240 Vac, 50/60 Hz
Operating Environment: 15ºC to 30ºC (59 to 86ºF)
Dimensions: 15.4"W x 16.3"D x 18"D (39.1W x 41.5D x 45.7H cm)
Weight: 55 lbs. (25 kgs) |