Helena catalog

ICHOR® Hematology Analyzer

Cat. No.

Item

ICH001

Ichor Hematology Analyzer, 110/220V


The ICHOR hematology instrument is a self-contained, compact, easy-to-use, automated impedance cell counter. When used with Plateletworks
reagents, the ICHOR system provides a hematology profile of a standard CBC with platelet count plus platelet aggregation or inhibition within five minutes. The test is simple. A fresh whole blood sample is added to both an EDTA baseline and a Collagen, ADP or Arachidonic Acid agonist tube. All tubes are run on the Ichor hematology analyzer. CBC, including platelet count, is available within minutes. Platelet aggregation is then calculated from results.


ICHOR / Plateletworks Advantages

CBC & Platelet Aggregation at Point of Care - provides immediate platelet count and aggregation, WBC, RBC, hemoglobin and hematocrit, RBC indices (MCV, MCH, MCHC), RBC, MPV and platelet histograms
Simple, rapid procedure - delivers critical hematology parameters and platelet aggregation in five minutes at the point of patient care
No special sample preparation - performed on whole blood using preloaded baseline and agonist tubes
Traditional cell counting principle - uses the laboratory standard of electronic impedance variation generated by the actual passage and counting of each cell type through the instrument’s microaperture
Traditional laboratory agonists - Plateletworks uses collagen, ADP and arachidonic acid, the same agonists used to perform platelet aggregation in the traditional laboratory setting
Automatic, closed tube sampling - optimizes safety and ease of use for any operator
Self-contained reagent pack and waste container - eliminates cumbersome external tubing sets, reagent containers and waste reservoirs, making technology suitable for point of care use
QC - easy to run in minutes with results stored for analysis and display on SmartCard (no manual charting)


ICHOR Specifications

Tests: Hemoglobin, Hematocrit, RBC Count, RBC Indices (MCV, MCH, MCHC), WBC Count, Platelet Count, MVP, Whole Blood Platelet Aggregation with ADP, Collagen, Arachidonic Acid

Reagent Delivery System: All-in-one reagent/waste pack contains reagent and self-included waste chamber

Linearity: RBC (x106 cells/µL): 0.2 to 7.5 ±0.05 (or ±3%)

                  WBC (x103 cells/µL): 0.5 to 80 ±0.2 (or ±3%)

                  Hemoglobin (g/dL): 2.5 to 23 ±0.3 (or ±2%)

                  Platelets (x103 cells/µL): 10 to 1,000 ±10 (or ±6%)
                  Hematocrit (%): 11.6 to 55 ±2 (or ±3%)
..................... and 56 to 62.4 ±5 (or ±5%)

Sample: Automated, closed-tube sampling requires only 10-12 µL of whole blood

Throughput: 70 seconds per sample; 60 samples per hour

Operating Mode: Sequential, single channel analysis

Operating Principle: Electronic impedance measurement

Special Features: Programmable high and low limits for each parameter. Programmable automatic cleaning cycle. Push button “standby” and “shut down” cycle. Calibration SmartCard for automatic instrument calibration.

Quality Control Features: Integrated QC program for data management and automatic report generation.

User Interface: Alphanumeric keypad

Display: 2 x 40 character backlit LCD

Printer: 4.5" (112 mm) thermal. (Dot matrix option also available.)

Printout: Includes test results, RBC and platelet histograms (if desired), high and low limit indications, alphanumeric patient identification, operator ID, time and date.

Memory: Stores last result in memory. Optional Memory SmartCards archive up to 60 patient data points per card.

Power Requirements: 90 to 240 Vac (self-adjusting), 50/60 Hz

Operating Environment: Room temperature

Dimensions: 16.5"H x 12.5"W x 14"D (42 x 32 x 36 cm)

Weight: 30.8 lbs. (14 kgs)